Most of the lab testing a medical assistant does in an office is CLIA waived: a glucose meter, a urine dipstick, a rapid strep test, a urine pregnancy test. The NHA CCMA exam puts this in Point of Care Testing and Laboratory Procedures, 9 of 150 scored items, and the questions are less about the biology of each test than about the rules for running it. This post follows the CLIA, quality control and results chapters of unit 14 in the free NHA CCMA course, drawing on the CLIA regulations and the CDC's "Ready? Set? Test!" booklet. It is a study aid, not clinical guidance; the provider interprets results and facility policy governs how tests are run and reported.

What CLIA is and who runs it

CLIA is the Clinical Laboratory Improvement Amendments of 1988. It regulates all laboratory testing on humans in the United States, with the goal of accurate, reliable and timely results. Three federal agencies share the work:

The three complexity levels

CLIA sorts tests into three categories:

  1. Waived
  2. Moderate complexity, which includes provider-performed microscopy (PPM)
  3. High complexity

Moderate and high together are called "nonwaived" testing. Nonwaived labs are inspected and must meet proficiency testing, quality control, quality assessment and personnel standards; moderate and high differ only in the personnel requirements.

A test qualifies as waived if it meets at least one of three conditions: it is FDA cleared for home use, it is so simple and accurate that errors are negligible, or there is no reasonable risk of harm if it is done incorrectly.

Two defaults are worth memorizing. A test that has not been categorized, a lab-developed test, and a test modified from the manufacturer's instructions all default to high complexity. And "point of care" describes where a test is done, not its complexity; a point-of-care test is not automatically waived. Provider-performed microscopy, such as urine sediment under the microscope, is moderate complexity done by physicians, dentists or midlevel practitioners, not by an MA under a Certificate of Waiver.

The Certificate of Waiver and its two duties

A Certificate of Waiver (CoW) is issued by CMS to a site that performs only waived tests. Sites doing nonwaived testing hold a certificate of compliance or of accreditation instead. A waived site has two federal duties:

  1. Follow the manufacturer's instructions for performing the test.
  2. Meet the Certificate of Waiver requirements.

That first duty carries more weight than it looks. The CDC requires the most current version of the manufacturer's instructions, checked with each new lot and shipment, with the old version filed and staff told about changes. Studies show most problems at waived sites come from failing to follow and understand those instructions. A guide or color chart is a helper only; it never replaces the current package insert.

Modifying a waived test, or using it off label, makes it high complexity and it can no longer be done under a CoW. The CDC's example is running a waived glucose meter on a patient whose hematocrit or oxygen level is outside the range the manufacturer specifies. CMS does not routinely inspect waived sites, though it may after a complaint or when there is risk of patient harm, and CLIA does not require proficiency testing for waived tests.

Common waived tests

The regulation itself lists a core set: nonautomated urine dipstick tests, nonautomated fecal occult blood, visual ovulation (LH) and urine pregnancy tests, nonautomated ESR, blood glucose on FDA-cleared home-use monitors, spun microhematocrit, and single-analyte hemoglobin instruments with a direct readout.

As of 2026, the FDA's waived analyte list is much longer and includes hemoglobin A1c, group A strep, influenza A and B, SARS-CoV-2, infectious mononucleosis antibodies, cholesterol, HDL and triglycerides, HIV antibodies, urine hCG, glucose, hemoglobin, hematocrit, ESR, fecal occult blood, prothrombin time, H. pylori, RSV and some drugs of abuse. Waived status belongs to a specific test system (brand and model), not to the analyte in general; only certain A1c or PT devices are waived.

Waived does not mean harmless. Prothrombin time and glucose are waived tests used to adjust anticoagulant and diabetes medication, so an error can hurt a patient. Specimen collection overlaps with phlebotomy, a separate CCMA subdomain; tube choice and sequence are covered in phlebotomy tube colors and additives and the order of draw, explained, and the blood draw tube color guide is a quick lookup.

Quality control

Quality control checks that a test performs as expected and warns of reagent deterioration, equipment failure, bad storage conditions or human error.

Internal (built-in) controls show the test is working: enough sample was added, it moved through the strip, the electronics work. The control line on a rapid strep or hCG cassette is the everyday example.

External controls are liquid or other materials that mimic patient samples and check the whole process from sample application to reading the result. A negative control should read negative and a positive control positive; quantitative tests use low and high (level 1 and level 2) controls with expected ranges. External controls are run exactly as a patient sample would be, by the same staff who test patients.

When to run QC:

If QC fails, patient results are not reported. Troubleshoot, repeat QC, and only when it passes retest the patient and report. The QC log records tester initials, date, test and control lot numbers, results and comments, with supervisor review.

Reagent rules from the same source: never mix components from different kit lots, date a vial or kit when it is opened and discard anything expired, and let refrigerated reagents and samples reach room temperature before testing. Refrigerator, freezer and room temperatures are logged daily; a missed reading is never filled in later, and an out-of-range temperature means stop testing and tell the supervisor.

Timing errors and the three phases

Timing is the classic waived-test error. Reading too soon can give an invalid or false negative result; reading too late can give a false positive as color overdevelops, a false negative as color fades, or an invalid result. Use a timer and follow the manufacturer's steps in order. Each urine dipstick pad has its own read time on the bottle label.

Testing errors are sorted into three phases: preanalytical (order, patient identification, collection, labeling, transport), analytical (performing the test, QC, calibration) and postanalytical (recording, reporting, critical value calls). Most errors are preanalytical, and hemolysis is the most common one and the leading cause of sample rejection.

If a result is invalid, does not fit the patient's clinical picture, or is outside the instrument's reportable range, the test is repeated, sometimes on a new sample. Record the first result, fix the problem, retest, and report only the final acceptable result.

In-range, out-of-range and critical values

Reference ranges vary between laboratories, and the lab's own reference interval must be available to the ordering provider, the same caveat that applies in NCLEX lab values: what to know for the exam. A few CDC cutoffs are common enough to know outright, per the current guideline:

TestNormalPrediabetesDiabetes
Fasting glucose99 mg/dL or below100-125 mg/dL126 mg/dL or above
A1CBelow 5.7%5.7% to 6.4%6.5% or above

A fasting test means nothing to eat for at least 8 hours. For urine, specific gravity runs about 1.005 to 1.030 and pH about 4.6 to 8.0; glucose, ketones, protein, bilirubin, nitrites and blood cells are not normally present. A positive nitrite suggests a urinary tract infection, but a negative one does not rule it out.

Recording rules: qualitative results are written in words or abbreviations ("Positive" or "Pos", "Negative" or "Neg"), never as "+" or "-" symbols. Quantitative results carry the manufacturer's units.

A critical value is a result that requires immediate notification of the clinician, and CLIA requires the lab to alert the ordering person immediately when a result shows an imminently life-threatening condition. Each site defines its critical values and documents when and to whom each one was reported. The MA's part is to notify the provider promptly per office policy and document it; the MA does not interpret the result or give the patient a diagnosis. Verbal reports are documented and followed by a written report, and results go only to authorized people, which is a HIPAA rule covered in HIPAA and patient consent for medical assistants.

How to study this subdomain

The rules above are process knowledge, and the exam asks them as "what should the MA do next" items: QC failed, a dipstick was read late, a glucose came back far outside the range. Practice retrieving the action, not just recognizing the term; the testing effect is the reason that works. Chain of infection and medical asepsis covers the cleaning and disinfection side of the same room. Point of care testing is 6% of the exam, so size your time using what's on the NHA CCMA exam, then work the unit 14 chapters of the NHA CCMA flashcard course.